| Catalog Number |
121610 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Obtain Sample (2533)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Expiry date: 06/2021.
|
| |
|
Event or Problem Description
|
|
It was reported that during a biopsy procedure, after the first sample pass, the device allegedly failed to obtain a tissue sample.It was further reported that the tissue was allegedly caught between the stylet and cannula, making it difficult to remove from the patient.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: one monopty needle was returned for evaluation.The device was able to prime, fire, and obtain samples without issue; therefore, the investigation is unconfirmed for failure to obtain samples.Although there were no issues removing the needle during functional testing, the conditions of use could not be recreated; therefore, the investigation for difficult to remove the needle from the patient is inconclusive.The definitive root cause could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review:a review of the product instructions for use (ifu) procedural instructions, indications, warnings, precautions, cautions, possible complications and contraindications showed that the product labeling is adequate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a biopsy procedure, after the first sample pass, the device allegedly failed to obtain a tissue sample.It was further reported that the tissue was allegedly caught between the stylet and cannula, making it difficult to remove from the patient.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|