| Model Number |
8575 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 100% stenosed target lesion was located in the moderately tortuous and mildly calcified artery.A 6f runway guide catheter was selected for use.However, when the guiding catheter was tried to insert during acute myocardial infarction case, it was noted that the second curve got kinked/broken.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Mfr site address 2: no.20905 int.A.Col.,cd.Industrial, 22444.Device evaluated by mfr.: returned product consisted of a runway guide catheter.The device showed 1 kink.The kink was located at the strain relief.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 100% stenosed target lesion was located in the moderately tortuous and mildly calcified artery.A 6f runway guide catheter was selected for use.However, when the guiding catheter was tried to insert during acute myocardial infarction case, it was noted that the second curve got kinked/broken.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|