| Model Number |
S401 |
| Medical Device Problem Codes |
High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Event or Problem Description
|
|
Boston scientific received information that the configuration on the right atrial (ra) lead switched to unipolar as a result of high out of range impedance measurements.While in unipolar configuration, there were a lot of episodes of oversensing.The configuration was reprogrammed to bipolar and minute ventilation (mv) was programmed on.After that, mv chop was observed.As a result, mv was programmed off and the noise was resolved.No further changes were made to the system.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Additional Manufacturer Narrative
|
|
The product has been received for analysis.This report will be updated upon completion of analysis.
|
| |
|
Event or Problem Description
|
|
Subsequent information was received that this device was explanted and returned.No additional adverse patient effects were reported.
|
| |
|