| Catalog Number |
TGU343420 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
03/21/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.(b)(4).According to the conformable gore® tag® thoracic endoprosthesis instructions for use (ifu), the complications associated with the use of the gore® tag® thoracic endoprosthesis may include but are not limited to endoleak, aortic expansion, and reoperation.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016, this patient underwent a repair of a thoracoabdominal aortic aneurysm whereby two conformable gore® tag® thoracic endoprostheses were implanted in zone 4 of the thoracic aorta.Two stents were also reportedly implanted in the celiac and superior mesenteric arteries.Pre-treatment computed tomography angiography (cta) showed that the maximum diameter of aortic aneurysm/lesion was 66mm.The patient tolerated the initial procedure.From (b)(6) 2016 till (b)(6) 2016, the follow up computed tomography angiography (cta) showed that the maximum diameter of aortic aneurysm/lesion was 70mm.On (b)(6) 2018, the follow up cta determined that the maximum diameter of aortic aneurysm/lesion was 76mm.Additionally, on (b)(6) 2018, a proximal type i endoleak associated with a tgu343420/14653201 was identified.On (b)(6) 2018, the reintervention was performed whereby an additional aortic endograft was implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information: event date.
|
| |
|
Search Alerts/Recalls
|