Merge unity pacs is designed to display a message to the user when an exam report fails to upload to the image server.In this instance, there was no evidence that the physician received the message warning that the report did not upload.It was not possible to determine if the message displayed on the screen for the physician when the report was first read and saved.The merge healthcare unity investigation showed the report did not upload successfully.The exam was read, there was an upload error.The locally cached report was corrupt or missing when universal manager tried to upload it.To resolve the issue, the exam was re-read by the physician.No other review/investigation will be performed for cause.
|
|
Merge unity pacs is a medical image and information management system that is used for viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems on (b)(6) 2018 the customer reported that an exam was completed, however, there is no exam report.The unity technician's investigation showed the report did not upload successfully.The exam was read, there was an upload error--the locally cached report was corrupt or missing when universal manager tried to upload it: the exam was re-read by the physician.While images are available, a finalized report not being available for subsequent review by the patient's physician could potentially result in a delay in care that could lead to harm.However, the customer did not allege any harm, serious injury or death as a result of this issue.(b)(4).
|