• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RICHARD WOLF GMBH GRASPING FORCEP INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RICHARD WOLF GMBH GRASPING FORCEP INSERT Back to Search Results
Model Number 8393.292
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Device Embedded In Tissue or Plaque (3165)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Rwmic will follow up with the user facility and manufacturer to obtain additional and missing information.A follow up report will be submitted upon receipt of new or additional information.Rwmic considers this case open.
 
Event or Problem Description
On september 18, 2018, the user facility reported the following to richard wolf medical instruments (rwmic): the tip of the insert broke off in the patient.Will the device be returned? yes.Was the device being used during a procedure when the issue occurred? yes.Specifically, was the device being used on a patient when the issue occurred? yes.Was there any injury or illness to patient or other personnel due to issue? no.Did the issue cause a delay in the procedure being performed that put the patient at risk? no.Was there a similar back-up device available for use? yes.Was the scheduled procedure completed? yes.How was the patient anesthetized? unknown the other components of the modular system, the pistol shaped handle and the insulated sheath tube, are considered suspect devices and a report will be submitted for each device.Grasping forcep insert - product number 8393.292.Pistol shaped handle - product number 83930072 (mdr 1418479-2018-00033).Insulated sheath tube - product number 8393933 (mdr 1418479-2018-00034).
 
Event or Problem Description
User facility returned the device to rwmic on (b)(6) 2018 and the evaluation was completed on (b)(6) 2018.The device appeared to used and broken - the pull rod was broken distally at the transition to the joint part.The reported condition was verified using visual testing.The root cause was determined to be user handling issues, cause of the situation was due to a mechanical overload.Excessive force during gripping caused the distal end of the pull to break.No manufacturing defects were identified.
 
Additional Manufacturer Narrative
Device labeling was reviewed for patient code and device codes, see below: patient code: not applicable, no patient problem was reported.Device code: ifu was reviewed, see below: caution.Excessive force can lead to breakage of or damage to the products.Due to the small dimensions required, the products have only limited strength.Use these products only to grasp and ablate small, soft tissue portions or organs.For therapeutic applications we recommend having a backup instrument at hand.Make sure that no missing parts remain in the patient.Caution! be careful if products are damaged or incomplete! injuries of the patient, user and others are possible.Run through the checks before and after each use.Do not use the products if they are damaged, incomplete or have loose parts.Return damaged products together with any loose parts for repair.Do not attempt to do any repairs yourself.Check the instruments and accessories for the following: damage, surface changes (e.G.Corrosion), sharp edges not suitable for the application, loose or missing parts, rough surfaces.Caution.Damaged surface in the joint area of the forceps.The joint pin may loosen.Check for surface changes, such as fissures, at the joint pin, user facility was contacted 3 times in an effort to collect additional information but none could be provided.Rwmic considers this case open.The manufacture will be contacted to collect missing information.A follow up report will be submitted upon receipt of new or additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRASPING FORCEP INSERT
Common Device Name
GRASPING FORCEP INSERT
Manufacturer (Section D)
RICHARD WOLF GMBH
pforzheimer street 32
knittlingen, 75438
GM  75438
MDR Report Key7982482
Report Number1418479-2018-00032
Device Sequence Number13887902
Product Code GCJ
UDI-Device Identifier04055207015988
UDI-Public04055207015988
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/18/2018
Report Date (Section F) 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8393.292
Device Catalogue Number8393.292
Device Lot Number51010295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date11/21/2018
Event Location Other
Date Report to Manufacturer12/21/2018
Initial Date Received by Manufacturer 09/21/2018
Supplement Date Received by Manufacturer09/21/2018
Initial Report FDA Received Date10/18/2018
Supplement Report FDA Received Date12/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Date Report Sent to FDA12/21/2018
-
-