| Model Number |
SE-A111512 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no known reported patient involvement associated with the complained event.It is unknown when device malfunctioned.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is synthes sales consultant.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that when sales consultant was going through the inventory on (b)(6) 2018 it was identified that one (1) hammerlock implant kit and one (1) speedarc compression implant kit (non-implanted stock) still sterile packaged were broken.Devices are still packaged and not opened.There was no patient and procedure involvement.This report is for one (1) speedarc compression implant kit 11x15x12mm.This is report 2 of 2 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device history: device history lot, part number: se-111512a, bme lot number: bmese140281c.Manufacturing date or release to warehouse date: 11 dec 2014.Place of manufacture: biomedical enterprises, (b)(4) lot expiration date: 5 nov 2019.Dhr review: a review of the device history record revealed there were complaint related anomalies during the manufacturing of this product.The device history record shows this lot of 11x15x12mm speed arc staples was processed through the normal manufacturing and inspection.Investigation summary complaint summary: it was reported on (b)(6) 2018, two (2) hammerlock implant kit were defective.It is unknown when the issue was discovered.It is unknown if there was patient and procedure involvement.Visual inspection was performed to determine the failure mode of this complaint.Inspection showed that the inserter broke, releasing the implant.Performed (b)(6) 2018.The items involved in this complaint are part of the bounding.Items were 100% compliant with bme requirements at the moment of release.After investigation, it was determined that the root cause was a molding issue from the provider (protolabs).No manufacturing related issue was identified and/or confirmed, therefore review to the specific prm and prm line is not required.The potential impact of the design in the complaint condition has previously been addressed.No further corrective action is necessary at this time.Conclusion: the complaint condition is confirmed as the speedarc compression implant kit (part # se-111512a, lot # bmese140281c) is broken.Replication of the complaint condition is not applicable as the inserter is already broken.A root cause could not be definitively determined.No manufacturing related issue was identified and/or confirmed.Thus, all further investigations and actions, including assessment of risk, corrective and/or preventative action is proposed or already launched.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.
|
| |
|
Event or Problem Description
|
|
Update,10.12.2018: the devices involved in this complaint were received by bme on oct.12, 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.
|
| |
|
Search Alerts/Recalls
|