| Model Number |
MMT-722NAS |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
09/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that they experienced high blood glucose level of 500 mg/dl.The customer did not provide details regarding symptoms or treatment of their high blood glucose episodes.The customer also reported that the buttons required multiple presses to work, unexplained no delivery alarm and motor rewind issue.The customer declined to troubleshooting for high blood glucose level.The product was not returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with all buttons responding properly.No damage or moisture damage on keypad assembly.No unlocked lcd keypad connector.However, device alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test and excessive no delivery test or verify no excessive no delivery/occlusion alarm due to motor error alarms.
|
| |
|