Model Number MMT-722NAS
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Customer reported via phone call that they experienced high blood glucose level of 500 mg/dl.The customer did not provide details regarding symptoms or treatment of their high blood glucose episodes.The customer also reported that the buttons required multiple presses to work, unexplained no delivery alarm and motor rewind issue.The customer declined to troubleshooting for high blood glucose level.The product was not returned for analysis.
 
Additional Manufacturer Narrative
Device received with all buttons responding properly.No damage or moisture damage on keypad assembly.No unlocked lcd keypad connector.However, device alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test and excessive no delivery test or verify no excessive no delivery/occlusion alarm due to motor error alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722NAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key7985361
Report Number2032227-2018-45965
Device Sequence Number16492637
Product Code OYC
UDI-Device Identifier00613994223746
UDI-Public(01)00613994223746
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-722NAS
Device Catalogue NumberMMT-722NAS
Device Lot NumberA000093830
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2018
Initial Date Received by Manufacturer 09/21/2018
Supplement Date Received by Manufacturer10/30/2019
Initial Report FDA Received Date10/19/2018
Supplement Report FDA Received Date11/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNOMED INF SETFRN-UNK-RSVR
Outcome Attributed to Adverse Event Other;
Patient Age44 YR
Patient Weight200
-
-