• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PFCSIGMA DIST FEMAUG TRL 4MMS2; KNEE INSTRUMENT : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PFCSIGMA DIST FEMAUG TRL 4MMS2; KNEE INSTRUMENT : FEMORAL TRIALS Back to Search Results
Catalog Number 961820
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the post broke off of the back of the augment trial.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the instrument associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFCSIGMA DIST FEMAUG TRL 4MMS2
Common Device Name
KNEE INSTRUMENT : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7985389
Report Number1818910-2018-72996
Device Sequence Number11121286
Product Code LXH
UDI-Device Identifier10603295240198
UDI-Public10603295240198
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number961820
Device Lot NumberGM3615801
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/16/2018
Supplement Date Received by Manufacturer11/28/2018
Initial Report FDA Received Date10/19/2018
Supplement Report FDA Received Date11/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-