BARD ACCESS SYSTEMS POWERPICC SV CATHETER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
N/A |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned, at this time, to the manufacturer for evaluation.A lot history review (lhr) of rebq2134 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the that picc line was leaking and the line needed to be exchanged or removed.Per notes in chart: (b)(6) 2018 vat called because dressing non-occlusive.Dressing changed and catheter found to be out 5 cm.Line flushed by bedside rn and catheter noted to be leaking.After speaking with providers line left in place (but not used) so a decision could be made to exchange of remove the line.After id consulted and iv antibiotics changed to po, the line was removed the next day.Almost a complete break in catheter noted at the point where the catheter joins to the wings.
|
| |
|