Catalog Number 71343203
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Injury (2348)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implant was found broken on the patient.
 
Additional Manufacturer Narrative
The affected oxinium femoral head was not returned for evaluation.Our investigation including a review of complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.Our clinical evaluation noted that no clinical relevant documents were provided to conduct a thorough medical assessment.No further medical assessment is warranted at this time.Without the return of the actual product involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXINIUM FEM HD 12/14 32MM -3
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7986790
Report Number1020279-2018-02238
Device Sequence Number1551993
Product Code LZO
UDI-Device Identifier03596010474216
UDI-Public03596010474216
Combination Product (Y/N)N
PMA/510(K) Number
K021673
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/20/2022
Device Catalogue Number71343203
Device Lot Number12AM12993
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/25/2018
Supplement Date Received by Manufacturer09/25/2018
Initial Report FDA Received Date10/19/2018
Supplement Report FDA Received Date01/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-