Additional Manufacturer Narrative
|
|
Date of event: unknown.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd¿ 10 ml syringe with interlink® vial access contained foreign matter.No serious injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: one opened blisterpack containing a 10ml syringe with needle attached was received in an envelope.The package was confirmed to be from batch #8010900 (p/n 303405).Stopper was observed to have a large piece of foreign matter attached to the top, in the fluid path.The foreign matter appeared to be a large piece of tape, larger than level 3 in size.Potential root cause for the foreign matter defect is associated with the assembly process.No corrective actions are necessary based on the defective rate identified.A device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.
|
| |
|
Event or Problem Description
|
|
It was reported that bd¿ 10 ml syringe with interlink® vial access contained foreign matter.No serious injury or medical intervention was reported.
|
| |
|