| Catalog Number |
6125127090 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Awaiting device evaluation.A follow up report will be filed once the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee procedure, the blade broke at the mount.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that the markings above the full insert line of the blade along with the heavy markings below the full insert text suggest that the handpiece was pushed forward while the blade was in use.The impaction marks on the edge of the blade coupled with the fact that the teeth were worn down suggests that the blade came into contact with metal while cutting.These factors could have contributed to the blade breaking.The ifu of handpieces associated with this device warn the user against this type of use: "do not apply excessive pressure, such as bending or prying, with the blade.Excessive pressure may bend or fracture the blade and result in tissue damage, loss of tactile control, and/or the ejection of blade fragments at a high velocity".The associated devices ifu's state that the device and associated handpieces are "intended for use in the cutting, drilling, decorticating, and smoothing of bone and other bone related tissue in a variety of surgical procedures." the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee procedure, the blade broke at the mount.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.
|
| |
|
Search Alerts/Recalls
|