MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B1120-060
Medical Device Problem Codes Off-Label Use (1494); Material Rupture (1546); Material Separation (1562); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).Evaluation summary: visual analysis was performed on the returned device.The reported balloon rupture and separation was confirmed.It should be noted that the armada 35 instruction for use warns: inflation in excess of the rated burst pressure may cause the balloon to rupture.Additionally, it was reported that the armada 35 was being used to treat the cephalic vein.It should be noted that the ifu states: the device is intended for dilatation of lesions in the renal, iliac, femoral, popliteal, tibial, and peroneal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.This device is also indicated for stent post-dilatation in the peripheral vasculature.It could not be determined if using off-label contributed to the reported difficulties.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined that the reported difficulties were likely due to case circumstances.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that during a procedure of the severely stenosed cephalic vein using a radial venous access, the 12 x 60 mm armada balloon dilatation catheter (bdc) was being inflated.During the first inflation above rated burst pressure (rbp) the balloon ruptured and detached.Several attempts using different access sites and sheath changes were unsuccessful; eventually, the balloon fragment was snared and removed from the vein.Although the procedure time was extended by approximately 2 hours, there was no clinical significance to the patient who remained stable throughout the procedure.There was no reported adverse patient sequela.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7986994
Report Number2024168-2018-08053
Device Sequence Number1409416
Product Code LIT
UDI-Device Identifier08717648154997
UDI-Public08717648154997
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Catalogue NumberB1120-060
Device Lot Number60409G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2018
Initial Report FDA Received Date10/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-