Method: visual inspection, product history review, compliant history review, nc/capa history review, labelling review, risk assessment result: the sales rep reported event was confirmed via visual inspection.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Conclusion: the potential causes of the reported event includes guide not properly attached to plate, guide/awl limits visibility, user tilts guide off axis affecting screw path, over-angulation of screwdriver which leads to unintended cantilever force, however a definite root cause cannot be determined at this time.
|