| Catalog Number |
48328500 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
09/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a screwdriver tip broke while inserting a screw.The tip was retrieved from the head of the screw and an alternate screwdrivers was used to fully seat the screw.The event resulted in 5 minute surgical delay to retrieve the screwdriver tip.No adverse consequences to the patient were reported.This complaint will be reported as a malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed via visual inspection of the returned device.Per visual inspection wear and deformation was identified around broken tip.The device is in the field for over 7 years.Manufacturing files were reviewed and no revealed anomalies were found.The most likely root cause of the tip fracture ifs normal wear due to device age.
|
| |
|
Event or Problem Description
|
|
It was reported that a screwdriver tip broke while inserting a screw.The tip was retrieved from the head of the screw and an alternate screwdrivers was used to fully seat the screw.The event resulted in 5 minute surgical delay to retrieve the screwdriver tip.No adverse consequences to the patient were reported.
|
| |
|