STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 12MM X 45CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0036121240
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device is unavailable.
 
Event or Problem Description
It was reported that the subject device prematurely deployed inside the catheter.Per physician¿s opinion may have been caused by friction inside the catheter.Additionally,the friction was observed while advancing the subject device through stent and the subject device deployed when it was pulled back.The subject device and the microcatheter were safely removed from the patient as one unit and there was no clinical consequence to the patient due to the reported event.
 
Additional Manufacturer Narrative
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.The subject device was not returned; therefore, physical as well as a functional evaluation could not be performed.Based on the information currently available an assignable cause for this event cannot be determined.
 
Event or Problem Description
It was reported that the subject device prematurely deployed inside the catheter.Per physician¿s opinion may have been caused by friction inside the catheter.Additionally,the friction was observed while advancing the subject device through stent and the subject device deployed when it was pulled back.The subject device and the microcatheter were safely removed from the patient as one unit and there was no clinical consequence to the patient due to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 12MM X 45CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7987197
Report Number3008881809-2018-00458
Device Sequence Number165604
Product Code HCG
UDI-Device Identifier04546540711533
UDI-Public04546540711533
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Catalogue NumberM0036121240
Device Lot Number20140363
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/27/2018
Supplement Date Received by Manufacturer11/05/2018
Initial Report FDA Received Date10/19/2018
Supplement Report FDA Received Date11/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATLAS STENT (STRYKER); MICOCATHETER (PHENOM)
-
-