| Model Number |
760 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
09/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the field service engineer evaluated the ventilator and replaced the erasable programmable read-only memory.The ventilator passed all testing per manufacturer specifications and was returned to the customer for clinical use.If the replaced part is returned for failure investigation, a supplemental medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during use on patient the 760 ventilator appears to have run a background check that generated both a constant audio and visual alarm.The ventilator stopped delivering breaths to the patient and safety valve opened.The ventilator went inoperative and also displayed post t del (power on self test) (delivered tidal volume) error messages.The patient was removed from the ventilator and placed on an alternate ventilator with no harm or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the erasable programmable read-only memory (eproms) were returned for failure investigation.The parts were visually inspected and functionally tested with no anomalies observed.Conclusion: no fault found.There was no malfunction or product deficiency identified.If information is provided in the future, a supplemental report will be issued.
|
| |
|