| Model Number |
ZXT150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) /2017 and (b)(6) 2018.(b)(4).Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxt150 19.5 diopter lens was implanted in the patient's right eye (od) on (b)(6) 2017.It was later explanted on (b)(6) 2018, due to the patient experiencing significant and troubling glare.There was no incision enlargement, no vitrectomy, and no sutures used.The replacement lens was a non-johnson and johnson lens.Reportedly, there was no patient injury.No additional information was provided.The patient had bilateral lens implants.This report will capture the lens implanted in the patient's right eye.A separate report will be filed for the left eye.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|
Search Alerts/Recalls
|