• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION TURBO-ELITE LASER ATHERECTOMY CATHETER; TURBO ELITE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION TURBO-ELITE LASER ATHERECTOMY CATHETER; TURBO ELITE Back to Search Results
Model Number 420-159
Medical Device Problem Codes Loss of or Failure to Bond (1068); Material Separation (1562); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Iatrogenic Source (2498); Device Embedded In Tissue or Plaque (3165)
Date of Event 09/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient gender, age, and weight were requested, but unavailable from the facility.Upon investigation of suspect catheter's distal tip, no marker band was present.Under microscopic evaluation a void was seen on the side wall of epoxy.Multiple fibers could be seen protruding out of the tip of epoxy.Evaluation confirmed reported issue.
 
Event or Problem Description
A philips representative reported that during an atherectomy procedure to remove a de novo lesion from the superficial femoral artery (sfa), the spectranetics turbo-elite laser catheter 420-159 was utilized.The turbo-elite laser catheter was advanced over the 0.014¿ guidewire, through a 6f sheath, from a pedal approach.The laser catheter advanced without issue, but the physician noted that the tip of the laser catheter (the radiopaque marker) was moving back and forth under x-ray, while he kept the laser catheter stationary.When he removed the laser catheter from the patient it was noticed that the radiopaque marker was not attached to tip of the laser catheter, and was still on the guidewire within the patient.The physician elected to stent across the radiopaque marker to the vessel wall.The radiopaque marker was then fixed in the patient¿s superficial femoral artery (sfa), and remained there post procedure.No further intervention was implemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-ELITE LASER ATHERECTOMY CATHETER
Common Device Name
TURBO ELITE
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
MDR Report Key7987513
Report Number1721279-2018-00149
Device Sequence Number11189295
Product Code MCW
UDI-Device Identifier00813132024765
UDI-Public00813132024765
Combination Product (Y/N)Y
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K170059
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/26/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/25/2020
Device Model Number420-159
Device Catalogue Number420-159
Device Lot NumberFBH18G18A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2018
Initial Report FDA Received Date10/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Date Report Sent to FDA01/01/2005
-
-