JOHNSON & JOHNSON SURGICAL VISION TECNIS 1-PIECE; MONOFOCAL IOLS Back to Search Results
Model Number ZCB00
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.If implanted, give date: not applicable, as the lens was not implanted.If explanted, give date: not applicable, as the lens was not implanted.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported when inserting zcb00 15.5 diopter intraocular lens (iol) into the eye the lens was scratched by the injector upon insertion.There was incision enlargement, no vitrectomy, and a similar lens was used to complete the procedure.Attempts were made to contact the customer account requesting additional information however, no response was received.No additional information was provided to johnson & johnson surgical vision.
 
Additional Manufacturer Narrative
Additional information: device available for evaluation: yes, returned to manufacturer on: 10/15/2018, device returned to manufacturer: yes.Device evaluation: only packaging components were received.The reported issue could not be verified.A product quality deficiency could not be determined.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no additional investigations request were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a malfunction or product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION
1700 e. st. andrew place
santa ana CA 92705
MDR Report Key7987628
Report Number2648035-2018-01361
Device Sequence Number16492706
Product Code HQL
UDI-Device Identifier05050474531628
UDI-Public(01)05050474531628(17)210815
Combination Product (Y/N)Y
Initial Reporter StateTX
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/15/2021
Device Model NumberZCB00
Device Catalogue NumberZCB0000155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 09/21/2018
Supplement Date Received by Manufacturer11/09/2018
10/25/2020
Initial Report FDA Received Date10/19/2018
Supplement Report FDA Received Date12/07/2018
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-