| Model Number |
ZA9003 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
03/23/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.If implanted, give date: unknown if lens was implanted, information not provided.If explanted, give date: unknown if lens was implanted and therefore unknown if explanted, information not provided.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that tecnis monofocal intraocular lens (model za9003 +26.0 diopter) had patient contact.Reportedly there was a broken capsule.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes, returned to manufacturer on: 11/20/2018, device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection using magnification was performed to the returned sample.Loose fibers/particles were observed on lens related the handling of the unit out of a sterile environment.Residues of viscoelastic material was observed on lens.Both haptics was also observed distorted.The condition in which the sample returned is consistent with a lens that was handled and prepare for surgical process.The complaint issue reported could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search of complaints was conducted and revealed three additional investigation requests for this production order number.No product deficiencies were identified.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
|
| |
|
Search Alerts/Recalls
|