| Model Number |
ZA9003 |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted; give date: not applicable as the product was not implanted.If explanted; give date: not applicable as the product was not implanted.Attempts were made to obtain missing information; however, to date no response was received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the haptic of za9003 13.0 diopter intraocular lens (iol) was torn off.No additional information was provided to johnson & johnson surgical vision.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation?: yes.Returned to manufacturer: 10/15/2018.Date received by manufacturer: 11/13/2018.Device returned to manufacturer: yes.Device evaluated by manufacturer: yes.Device evaluation: the product was received at manufacturing site.The lens is missing a portion (iol torn) including a haptic.The detached portion of the lens was not returned.Other cosmetic damages are observed on the lens edge.The reported code of haptic damage was not verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints was conducted, the search revealed no additional investigation requests have been received for this production order number.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product malfunction or product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|