The reported event is related to a l4-l5 procedure that was performed at (b)(6) medical center', us, on (b)(6) 2016.During the procedure, upon sending the rbt to l4 left trajectory, the surgeon did not change the rbt device's station on the platform according to the software instructions.Therefore, the trajectory was drilled from the wrong station which led to inaccuracy.Consequently, the surgeon drilled l4 left trajectory inside the disc space.Flouro shot confirmed the reported finding.Surgeon increased dissection and used bovie to contain the bleeding.Due to drop in blood pressure the surgeon called in a vascular surgeon.Finally, the case was aborted due to an excessive drop in blood pressure and the patient was brought back to complete the surgery the following day.The procedure was performed with the renaissance system successfully with no further complications.There was no residual damage to the patient from the incident.
|