| Model Number |
012001 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic living donor nephrectomy, the rubber part at the tip has peeled off.The procedure was completed with another device.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The jaw operated properly upon actuation of the handle but failed a grasping test.The device was disassembled and damage to the probe housing grove was observed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the secondary condition of damage to the probe housing groove is consistent with excessive material or an obstruction being compressed in the jaws as the trigger is connected to the groove and the jaws are closed.Any damage to the groove interface will result in the handle bottoming out and the jaws not closing completely.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|