Catalog Number 625-0T-36H
Medical Device Problem Codes Crack (1135); Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 09/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding impingement (rom), crack/fracture and pain involving a trident liner was reported.Crack/fracture of the liner was confirmed on review of the provided photo.Rom and pain was not confirmed.Method & results: device evaluation and results: the device was not returned however a photo was provided.A visual inspection of this photo confirmed crack/fracture of the ceramic element of the liner.The device was otherwise unremarkable.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: all devices in the reported lot were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar relevant events for the reported lot.Conclusions: crack/fracture of the ceramic head was confirmed on review of the provided photo however the root cause could not be determined.Rom and pain was not confirmed.Additional information, including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available and/ or the product is returned, this investigation will be re-opened.Device not returned.
 
Event or Problem Description
Dr.Stated that the cup had shifted causing an impingement on the rim of the ceramic cup which ultimately resulted in a fracture of the ceramic.Dr.Removed the ceramic liner and ceramic head.After debridement he replaced with a new ceramic liner and ceramic head.He placed a v 4toc taper sleeve resulting in a fresh trunnion.Item not explanted: 6021-0537.Lot 8509002.Update 02/october/2018: as reported in how was issue noticed: patient presented with pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT ALUMINA INSERT
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNC
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7990761
Report Number0002249697-2018-03450
Device Sequence Number1572180
Product Code MRA
UDI-Device Identifier04546540516879
UDI-Public04546540516879
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P000013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2010
Device Catalogue Number625-0T-36H
Device Lot Number12302102
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2018
Initial Report FDA Received Date10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age78 YR
Patient Weight135
-
-