Model Number CIE3-11-28-00
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Pain (1994)
Date of Event 09/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the history of the device, a potential contributing factor to the device movement is the endplate was under prepared evidenced by osteophytes or bone spurs on the posterior end of the vertebral bodies of the patient (based on reporting) coupled with device used was potentially undersized evidenced by the surgeon removing the device in an expanded state, these factors have contributed to device migration in the past; however without images, it is difficult to confirm the device moved post-op (versus not placed correctly at index) or to determine the exact cause of any movement that might have occurred.However, a device malfunction or failure is not likely to have occurred based on the device history review and the information known from the removal surgery (device removed intact, expanded, and full of bone).
 
Event or Problem Description
Varilift-lx removal surgery 1 month post-op.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VARILIFT-LX
Common Device Name
LUMBAR INTERBODY FUSION DEVICE
Manufacturer (Section D)
WENZEL SPINE, INC.
1130 rutherford lane
suite 200
austin TX 78753
MDR Report Key7990869
Report Number3008009850-2018-00006
Device Sequence Number1681960
Product Code MAX
UDI-Device Identifier00813210020139
UDI-Public00813210020139
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K151900
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/09/2023
Device Model NumberCIE3-11-28-00
Device Lot NumberBJ008E
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2018
Initial Report FDA Received Date10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;