• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH SCORPIO NRG X3 PS TIBIAL INSERT #11 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH SCORPIO NRG X3 PS TIBIAL INSERT #11 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 82-7-1115
Medical Device Problem Codes Fracture (1260); Unstable (1667)
Health Effect - Clinical Code Injury (2348)
Date of Event 09/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Left knee poly exchange.Update 02/october/2018: as reported in how was issue noticed: "approximately 10 years after implant.Patient has positive posterior instability".Explant picture provided by the rep.
 
Event or Problem Description
Left knee poly exchange.Update 02/october/2018: as reported in how was issue noticed: "approximately 10 years after implant.Patient has positive posterior instability".Explant picture provided by the rep.
 
Additional Manufacturer Narrative
Reported event: an event regarding instability and crack/fracture involving a scorpio tibial insert was reported.Medical review confirmed the event but could not determine a root cause.Method & results: -product evaluation and results: the device was not returned however one photo was provided of the explanted device.The photo shows a fracture of stabilizing ps post.It also shows some biological material on the device.-medical records received and evaluation: a review of the provided x-rays and explant photos by a clinician noted the following: it was reported that approximately 10 years after tka implantation that the patient developed posterior instability requiring revision surgery.A single ap and lateral x-ray of the knee was available for review.These x-rays demonstrate a cemented left cruciate retaining tka in anatomic position.The implants are in normal alignment and appear well fixed without evidence of loosening or mechanical complication.A photograph of the explanted ps polyethylene liner shows that the stabilizing ps post had fractured.-product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.-complaint history review: there were no other events for the lot provided.Conclusions: the medical review concluded the following: the event of posterior instability requiring revision surgery of a tka implanted in 2009 is confirmed.The cause of the posterior instability is fracture of the stabilizing ps post of a ps polyethylene tibial insert.The root cause of the fracture of the post of the insert is not identified.Documentation which would aid in the further completion of this assessment would include: ¿ operative reports ¿ dated pre and post op x-rays/ imaging from the index and revision surgeries ¿ outpatient office/clinic notes ¿ implant retrieval with material analysis no further investigation for this event is possible at this time as no devices and insufficient information was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCORPIO NRG X3 PS TIBIAL INSERT #11 15MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7991287
Report Number0002249697-2018-03454
Device Sequence Number13896616
Product Code JWH
UDI-Device Identifier07613327031256
UDI-Public07613327031256
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K071991
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2013
Device Catalogue Number82-7-1115
Device Lot NumberPRTMDE
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/24/2018
Supplement Date Received by Manufacturer12/21/2018
Initial Report FDA Received Date10/22/2018
Supplement Report FDA Received Date01/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
Patient Weight100
-
-