| Catalog Number |
0375627000 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the base plastic component was burned by the heat of the friction at the bur shaft and there were no burns to the patient or users.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: a fire broke out at the proximal end of the shaver.It happened during use inside the joint, at 9000 rpm.The failure identified in the investigation is consistent with the complaint record.The probable root cause/s could be excessive force applied by the user with poor or no suction during use.The excessive force may cause friction due to the inner shaft assembly and housing assembly interaction.Another possibility the cutter was clogged or inadequate suction would prevent fluid being removed to provide cooling to the shaft.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the base plastic component was burned by the heat of the friction at the bur shaft and there were no burns to the patient or users.
|
| |
|