Model Number 5537-G-511
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reported event: an event regarding seating/locking issues involving a triathlon insert was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as the reported device was not returned for evaluation.Medical records received and evaluation: a medical review was not performed because no medical information was provided.Device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: review indicated there have been no other similar events for the reported lot.Conclusions: the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation and insufficient information was provided.Further information such as return of device and surgical notes are needed to investigate this event further.If additional information and/or device becomes available, this investigation will be reopened.
 
Event or Problem Description
During a revision of patient's left knee, a 5x11 ts insert was implanted and removed because the locking mechanism was compromised.Rep reported there was no surgical delay.
 
Additional Manufacturer Narrative
Upon review, h1 was noted incorrect.This report is to correct h1.H3 other text : device not returned.
 
Event or Problem Description
During a revision of patient's left knee, a 5x11 ts insert was implanted and removed because the locking mechanism was compromised.Rep reported there was no surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NO 5. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7991655
Report Number0002249697-2018-03461
Device Sequence Number16564527
Product Code MBH
UDI-Device Identifier07613327027297
UDI-Public07613327027297
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date04/30/2023
Device Model Number5537-G-511
Device Catalogue Number5537-G-511
Device Lot NumberW515M7
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/24/2018
Supplement Date Received by Manufacturer12/23/2020
Initial Report FDA Received Date10/22/2018
Supplement Report FDA Received Date12/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age80 YR
-
-