| Model Number |
5537-G-511 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding seating/locking issues involving a triathlon insert was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as the reported device was not returned for evaluation.Medical records received and evaluation: a medical review was not performed because no medical information was provided.Device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: review indicated there have been no other similar events for the reported lot.Conclusions: the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation and insufficient information was provided.Further information such as return of device and surgical notes are needed to investigate this event further.If additional information and/or device becomes available, this investigation will be reopened.
|
| |
|
Event or Problem Description
|
|
During a revision of patient's left knee, a 5x11 ts insert was implanted and removed because the locking mechanism was compromised.Rep reported there was no surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon review, h1 was noted incorrect.This report is to correct h1.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
During a revision of patient's left knee, a 5x11 ts insert was implanted and removed because the locking mechanism was compromised.Rep reported there was no surgical delay.
|
| |
|