| Model Number |
43435 |
| Health Effect - Clinical Codes |
Hyperglycemia (1905); Hypoglycemia (1912); Overdose (1988)
|
| Date of Event |
10/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complainant stated she tested herself when she felt low and "was around 79mg/dl".This value could not be confirmed while on the phone with the customer as she was having difficulty manipulating the meter and visualizing the numbers.During the call to customer support, it was revealed that the consumer did not perform a control solution test for integrity before use their initial test strips as instructed in our directions for use.Per label copy/ package insert.High or low blood glucose results can indicate potentially serious medical conditions.In case of an unexpected result, you should repeat the test using a new test strip.If the result is still unexpected, or the reading is not consistent with how you feel, contact your hcp and treat as prescribed.Any change in the treatment of your diabetes should be discussed with your hcp.Nova max test strip insert- quality control.Checking the system control solution test: the nova max control solution is used as a quality control check to make sure that your blood glucose monitor and the nova max glucose test strips are working correctly.Do a control solution test: each time you open a new vial of test strips.Storage and handling.Keep the nova max glucose test strips vial tightly closed when not in use.Test strips should be stored only in the original vial.Meter and test strips are expected to be returned for evaluation.
|
| |
|
Event or Problem Description
|
|
The complainant stated that "because of her 209mg/dl reading she took an extra dose of geneva along with normal pill doses".She then reported that she went food shopping and felt "low" she ate a chocolate candy bar and felt better.
|
| |
|
Additional Manufacturer Narrative
|
|
The alleged deficiency could not be confirmed.Although the reported results were found in the memory of the returned device, failure analysis of the returned product revealed that the novamax plus glucose monitoring device performed within specification.The complainant returned the blood glucose test strips from the lot in question.Visual as well as chemical evaluation (by testing with control solution) was not able to replicate the alleged issue as the performance testing revealed the strips to perform within specification as well.The dhr was complete and contained all relevant data indicating the product put into finished goods met all specifications.Nova biomedical will continue to monitor as part of our post market surveillance program.
|
| |
|
Search Alerts/Recalls
|