| Model Number |
8888541023 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, there were leakage and fissure at the blue connector of the palindrome.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led a photographic evaluation of one device.A visual inspection of the returned photo noted: the catheter is in a sample bag, the cannula is detached from the hub and is not shown.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, there was a leakage and fissure at the blue connector of the catheter.No further information is available.
|
| |
|