Catalog Number 213701
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate evaluation statement: the complaint device was received and inspected.Visual observation reveals the thumb screw was broken and the femoral rod is stuck inside the guide frame, confirming this complaint.It is likely that the user hit the thumbscrew while malleting out the guide.Other than this possibility, we cannot discern a root cause for this failure.A device history record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).Associated medwatch report number: 1221934-2018-54772.
 
Event or Problem Description
It was reported by the sales rep via phone that during an acl repair procedure the sales rep's rigidfix femoral guide had the thumb screw break inside the guide.The break in the screw happened outside of the patient with no fragments created.The sales rep's 8.5 rigidfix femoral rod is jammed in the guide because the screw cannot be removed from the guide.The case was completed with this device with no patient harm or delays.The devices are being returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
BONENAIL GUIDEWIRE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7991974
Report Number1221934-2018-55013
Device Sequence Number202261
Product Code FZX
UDI-Device Identifier10886705002955
UDI-Public10886705002955
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number213701
Device Lot Number1701001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/22/2018
Initial Report FDA Received Date10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-