MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC HANDLE FOR 90° SCREWDRIVER; DRILL, BONE, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC HANDLE FOR 90° SCREWDRIVER; DRILL, BONE, POWERED Back to Search Results
Model Number 03.505.004
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/05/2018
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional product codes: dzj, hxx.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested.Device was used for treatment, not diagnosis.
 
Event Description
It was reported that on (b)(6) 2018 a patient underwent an unknown procedure wherein a 90 degree drill would not drill intra-operatively.There was no known delay on the surgery and no other medical intervention done.The procedure was completed successfully and the patient status is good.This complaint involves two (2) devices.This report is 1 of 2 for (b)(4).
 
Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device history records review was completed for part: 03.505.004, lot: 8165482.Manufacturing location: (b)(4).Release to warehouse date: mar 16, 2017.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDLE FOR 90° SCREWDRIVER
Type of Device
DRILL, BONE, POWERED
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7991997
MDR Text Key124678183
Report Number2939274-2018-54523
Device Sequence Number1
Product Code DZI
UDI-Device Identifier10887587013305
UDI-Public(01)10887587013305
Combination Product (y/n)N
PMA/PMN Number
K082649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03.505.004
Device Catalogue Number03.505.004
Device Lot Number8165482
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/05/2018
Initial Date FDA Received10/22/2018
Supplement Dates Manufacturer Received11/01/2018
Supplement Dates FDA Received11/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-