BOSTON SCIENTIFIC CORPORATION RADIAL JAW 4; FORCEPS, BIOPSY, NON-ELECTRIC Back to Search Results
Model Number M00513412
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a radial jaw 4 standard capacity biopsy forceps was used in the esophagus during an esophagogastroduodenoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, a foreign body was found inside the cup when they opened the biopsy forceps.The foreign body dropped into the patient and they were unable to find it.They used another of the same forceps to complete the case.Additionally, it was reported that the foreign body was light yellow orange that looked like the catheter of the forceps.There were no serious injury nor adverse patient effects reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Event or Problem Description
It was reported to boston scientific corporation that a radial jaw 4 standard capacity biopsy forceps was used in the esophagus during an esophagogastroduodenoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, a foreign body was found inside the cup when they opened the biopsy forceps.The foreign body dropped into the patient and they were unable to find it.They used another of the same forceps to complete the case.Additionally, it was reported that the foreign body was light yellow orange that looked like the catheter of the forceps.There were no serious injury nor adverse patient effects reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Additional Manufacturer Narrative
(b)(4).Visual inspection found the device does not have any visible failure or residues that indicates that a foreign matter was present.Based on the available information, the issue occurred during the procedure.As the device was withdrawn from the pouch and used, it is difficult to determine where the reported foreign matter was acquired.Therefore, the conclusion code selected for this complaint is cause not established.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly, and performance specifications at the time of release for distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIAL JAW 4
Common Device Name
FORCEPS, BIOPSY, NON-ELECTRIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key7992703
Report Number3005099803-2018-60976
Device Sequence Number1255805
Product Code FCL
UDI-Device Identifier08714729778004
UDI-Public08714729778004
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/13/2021
Device Model NumberM00513412
Device Catalogue Number51309
Device Lot Number0022375917
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Initial Date Received by Manufacturer 09/26/2018
Supplement Date Received by Manufacturer11/12/2018
Initial Report FDA Received Date10/22/2018
Supplement Report FDA Received Date12/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-