• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC AIMING ARM FOR SUPRAPATELLAR; NAIL,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC AIMING ARM FOR SUPRAPATELLAR; NAIL,FIXATION,BONE Back to Search Results
Model Number 03.010.441
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.The investigation summary: the investigation could not be completed; no conclusion could be drawn.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.
 
Event or Problem Description
It was reported that during a procedure on an unknown date, the drill bit was missing nails attached to a suprapatellar insertion handle.The unknown screw and drill bit was successfully inserted into the nail.No further incident but the initial part of the drill bit did not goes to the nail.It is unknown if there was a surgical delay.Patient and procedure outcome is unknown.Concomitant devices reported: unknown insertion handle (part # unknown, lot # unknown, quantity 1), unknown screw (part # unknown, lot # unknown, quantity 1), unknown nail (part # unknown, lot # unknown, quantity unknown).This complaint involves three (3) device.This report is 3 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during a procedure on (b)(6) 2018, while drilling through the aiming arm using the supra patellar insertion handle, the calibrated three-fluted drill bit missed the static transverse hole in the titanium cannulated tibial nail.It was meant to match up with the applied physical deformation to the insertion handle/aiming arm construct in order to achieve correct alignment which allowed the drill bit to pass through the nail as desired.It was further noted that the whole construct: insertion handle, aiming guide, and drill sleeves allow for enough play that certain patients' bone/tissue deform the construct enough to create misalignment.Following this, the titanium locking screw was inserted as intended without further incident.No further incident but the initial part of the drill bit did not go to the nail.It is unknown if there was a surgical delay.Patient and procedure outcome is unknown.Concomitant devices: titanium locking screw (part: 04.005.428s, lot: unknown, quantity: 1), titanium cannulated tibial nail (part: 04.034.349s, lot: unknown, quantity:1) this report is for one (1) aiming arm.This is report 3 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMING ARM FOR SUPRAPATELLAR
Common Device Name
NAIL,FIXATION,BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7992753
Report Number2939274-2018-54530
Device Sequence Number13888068
Product Code JDS
UDI-Device Identifier10886982068835
UDI-Public(01)10886982068835
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K111667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.010.441
Device Catalogue Number03.010.441
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/06/2018
Supplement Date Received by Manufacturer10/19/2018
Initial Report FDA Received Date10/22/2018
Supplement Report FDA Received Date11/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-