• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM Back to Search Results
Catalog Number 1012534-60
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.A query of the complaint handling database revealed no indication of a lot specific product issue.The investigation was unable to determine a cause for the reported inaccurate delivery.It may be possible that the distal sheath was restricted or entrapped within the anatomy such that when the sheath finally released and the stent deployed, the stent jumped due to the built-up tension within the shaft lumens of the delivery system causing a spring like release of the stent.However, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the proximal ostium of the superficial femoral artery.A 6.0x60mm absolute pro vascular self-expanding stent system (sess) was advanced to the lesion; however, during deployment the stent jumped about 5 millimeters.The stent was post-dilated with a 5x40mm armada 35 balloon catheter.Although the stent was in the target lesion an additional 6.0x20mm absolute pro vascular sess was advanced and deployed to cover the remainder of the target lesion.The procedure was successfully completed with post-dilation with an armada 35 balloon catheter.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7992958
Report Number2024168-2018-08094
Device Sequence Number11066284
Product Code NIO
UDI-Device Identifier08717648175992
UDI-Public08717648175992
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Catalogue Number1012534-60
Device Lot Number6100761
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2018
Initial Report FDA Received Date10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-