| Catalog Number |
140.662 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Pain (1994); Swelling (2091); Tissue Damage (2104)
|
| Date of Event |
12/21/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reported that 2 months after the dental implant was placed in ada site 5 in the mouth, the implant did not achieve osseo- integration.Clinician noted patient's pain and swelling.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.After the evaluation was noticed that the item received is different from the item reported.Therefore, the item was corrected and the lot number was not considered.Considering the surgical methodology, it was seen that the dentist has used less drills than recommended to perform the implant installation.
|
| |
|
Event or Problem Description
|
|
The clinician reported that 2 months after the dental implant was placed in ada site 5 in the mouth, the implant did not achieve osseo- integration.Clinician noted patient's pain and swelling.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.(b)(4).
|
| |
|
Search Alerts/Recalls
|