NEODENT - JJGC S.A. CM TITAMAX EX ACQUA IMPLANT 3.5X11; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 140.662
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Pain (1994); Swelling (2091); Tissue Damage (2104)
Date of Event 12/21/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
 
Event or Problem Description
The clinician reported that 2 months after the dental implant was placed in ada site 5 in the mouth, the implant did not achieve osseo- integration.Clinician noted patient's pain and swelling.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
 
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.After the evaluation was noticed that the item received is different from the item reported.Therefore, the item was corrected and the lot number was not considered.Considering the surgical methodology, it was seen that the dentist has used less drills than recommended to perform the implant installation.
 
Event or Problem Description
The clinician reported that 2 months after the dental implant was placed in ada site 5 in the mouth, the implant did not achieve osseo- integration.Clinician noted patient's pain and swelling.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM TITAMAX EX ACQUA IMPLANT 3.5X11
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key7993093
Report Number3008261720-2018-05099
Device Sequence Number13805785
Product Code DZE
UDI-Device Identifier07899878016670
UDI-Public07899878016670
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K133592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 10/22/2018
Report Date (Section F) 10/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/09/2018
Device Catalogue Number140.662
Device Lot Number66860
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/17/2018
Device Age1 MO
Event Location Other
Date Report to Manufacturer10/09/2018
Initial Date Received by Manufacturer 10/09/2018
Supplement Date Received by Manufacturer11/27/2018
Initial Report FDA Received Date10/22/2018
Supplement Report FDA Received Date12/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GRAFT; GRAFT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
-
-