SMITH & NEPHEW, INC. SYN POR FEM COMP SZ 15; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Model Number |
71306615 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Injury (2348)
|
| Date of Event |
04/26/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that right hip revision surgery was performed due to a fracture of femoral stem.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint devices were not returned.The clinical/medical team concluded, it was reported that right hip revision surgery was performed due to a fracture of the femoral stem.The revision operative report noted there was nothing to indicate what could have caused the fracture.The explants were sent to evaluation, but the results were not provided.Conclusion: it was noted during revision that there was nothing to indicate what could have caused the fracture of the femoral stem.The explanted stem was reported to have been sent to for evaluation; however, no results were provided for review.Without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the reported fractured femoral stem cannot be confirmed and it cannot be concluded that the reported event was associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.No further clinical assessment is warranted at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches/failure mode.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
The devices, used in treatment were not returned for evaluation, reporting event could not be confirmed.These devices were manufactured in 2007, 2008, 2009 and 2010.A clinical evaluation was conducted and it was noted during revision that there was nothing to indicate what could have caused the fracture of the femoral stem.The explanted stem was reported to have been sent to for evaluation; however, no results were provided for review.Without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the reported fractured femoral stem cannot be confirmed and it cannot be concluded that the reported event was associated with a mal-performance of the implant.A review of the device history records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of the risk management file and instructions for use document for this failure mode was conducted.Factors and/or potential causes that could contribute to the reported event have been identified in the risk management file includes but not limited to traumatic injury, lifetime of device, and overuse.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
|
| |
|
Search Alerts/Recalls
|
|
|