COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 980
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The service engineer (se) inspected the device, reviewed the logs and verified the reported issue.The se found a leak, delivery pressure solenoid (psol) stuck open error code.The se replaced the psol.The ventilator passed all testing per manufacturing specification and was placed back into clinical use.If the replaced part is returned for failure investigation, a supplemental medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, while in use on a patient a 980 ventilator activated backup ventilation.The patient was removed from the vent ilator and placed on an alternate ventilator with no harm or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
michael collins road mervue
galway
MDR Report Key7993534
Report Number8020893-2018-00517
Device Sequence Number1411590
Product Code CBK
UDI-Device Identifier10884521172876
UDI-Public10884521172876
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
K162738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number980
Device Catalogue Number980X3ENDRAC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2018
Initial Report FDA Received Date10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age35 YR
Patient Weight85
-
-