• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC PERFORMER INTRODUCER; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC PERFORMER INTRODUCER; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G11672
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) number = pre-amendment.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, the performer introducer was removed from the packaging in it's usual state with the dilator backloaded.During preparation of the device, the dilator was removed from the tip of the performer introducer and was inserted through the valve with difficulty but was fully inserted.Upon insertion into the patient's left common iliac artery, dilator removal was attempted but proved difficult.The dilator was eventually removed using hemostats and a vigorous pull which broke the hub end of the dilator.Another device was used to successfully complete the fenestrated endovascular aneurysm repair procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
Additional information regarding the patient and/or event has been received since the previous medwatch report was sent.This information is detailed.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: previously unreported additional information was received from the customer.The customer confirmed that the device was used over a.035 size wire guide, and that the patient's anatomy was slightly calcified.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation evaluation.A review of the complaint history, device history record, manufactures instructions, quality control, specifications, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned package confirmed that one rcfw-18.0p-38-30-rb was returned for investigation.The dilator was returned partially inserted into the sheath.There was no hub on the dilator and it was not returned with the device.The dilator was removed from the sheath with excessive force after the device was soaked with water.There appeared to be hemostat marks on the proximal end of the dilator.Biomatter was present in the valve of the device.The proximal fittings were disassembled for further examination, and in related pr252319, it was determined that there was glue partially occluding the lumen of the check-flo body.Additionally, a document based investigation evaluation was performed.There is evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should also be noted the only other complaints for this lot number were pr252319 & pr228045 which were created from this complaint for trending purposes.Furthermore, a review of the manufactures instructions and quality control procedures was conducted, and no gaps were discovered.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could be traced to manufacturing.There was evidence of excess glue applied when the sheath proximal fittings were attached, making this device out of specification.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORMER INTRODUCER
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7993713
Report Number1820334-2018-03109
Device Sequence Number19267332
Product Code DYB
UDI-Device Identifier00827002116726
UDI-Public(01)00827002116726(17)200831(10)8145376
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model NumberG11672
Device Catalogue NumberRCFW-18.0P-38-30-RB
Device Lot Number8145376
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2018
Initial Date Received by Manufacturer 10/09/2018
Supplement Date Received by Manufacturer11/14/2018
02/21/2019
Initial Report FDA Received Date10/22/2018
Supplement Report FDA Received Date12/12/2018
02/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-