It was reported, the user opened the package and removed the torcon nb advantage angiographic catheter when the head of the catheter fell.Off the tip was described as broken.This device did not make patient contact.The user changed to another torcon nb advantage angiographic catheter and successfully completed the procedure.No patient adverse effect was reported.Upon preliminary visual inspection by the manufacturer, it was noted the tip is broken off.The missing tip piece was not returned for investigation.
|
|
Investigation/evaluation: a review of the dimensional verification, complaint history, device history record, documentation, drawing, manufactures instructions, quality control, specification, and a visual inspection of the returned device was conducted during the investigation.The visual inspection of the returned device confirmed that biomatter was present on the proximal hub and throughout the shaft.A portion of the distal tip was missing.The surface appeared scraped and compressed at the distal tip on the shaft from 5cm to 5.5cm proximal to the distal tip, and 11.5cm to 12.5cm proximal to the distal tip.Length of distal tip after the bond site was 0.4cm.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, a review of the manufactures instructions and quality control procedures was conducted, and no gaps were discovered.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be established.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|