• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC TORCON NB ADVANTAGE ANGIOGRAPHIC CATHETER; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC TORCON NB ADVANTAGE ANGIOGRAPHIC CATHETER; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number G05955
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Non-healthcare professional.Pma/510(k) number = pre-amendment.Device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, the user opened the package and removed the torcon nb advantage angiographic catheter when the head of the catheter fell.Off the tip was described as broken.This device did not make patient contact.The user changed to another torcon nb advantage angiographic catheter and successfully completed the procedure.No patient adverse effect was reported.Upon preliminary visual inspection by the manufacturer, it was noted the tip is broken off.The missing tip piece was not returned for investigation.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation/evaluation: a review of the dimensional verification, complaint history, device history record, documentation, drawing, manufactures instructions, quality control, specification, and a visual inspection of the returned device was conducted during the investigation.The visual inspection of the returned device confirmed that biomatter was present on the proximal hub and throughout the shaft.A portion of the distal tip was missing.The surface appeared scraped and compressed at the distal tip on the shaft from 5cm to 5.5cm proximal to the distal tip, and 11.5cm to 12.5cm proximal to the distal tip.Length of distal tip after the bond site was 0.4cm.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, a review of the manufactures instructions and quality control procedures was conducted, and no gaps were discovered.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be established.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORCON NB ADVANTAGE ANGIOGRAPHIC CATHETER
Common Device Name
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7993740
Report Number1820334-2018-03220
Device Sequence Number1488067
Product Code DQO
UDI-Device Identifier00827002059559
UDI-Public(01)00827002059559(17)210511(10)8789427
Combination Product (Y/N)N
Initial Reporter CountryCN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/11/2021
Device Model NumberG05955
Device Catalogue NumberHNB5.0-38-80-P-NS-C2
Device Lot Number8789427
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2018
Initial Date Received by Manufacturer 08/17/2018
Supplement Date Received by Manufacturer01/02/2019
Initial Report FDA Received Date10/22/2018
Supplement Report FDA Received Date01/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-