• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETERIZATION TRAY WITH PLASTIC CATHETER; URETHRAL CATHETER TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETERIZATION TRAY WITH PLASTIC CATHETER; URETHRAL CATHETER TRAY Back to Search Results
Model Number 772514
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the catheters were missing from the catheter tray.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use was found adequate and states the following: "visually inspect the product for any imperfections or surface deterioration prior to use.1.Open csr wrap to form sterile field.2.Place underpad beneath patient, plastic side down.3.Put on cuffed gloves.4.Position drape on patient.5.Remove top tray.6.Lubricate catheter which is pre-attached to collection bag.7.Bag and catheter may be placed in basin until needed.8.Prep patient with povidone-iodine swabsticks.9.Proceed with catheterization in usual manner.10.Top tray may be placed on basin to help prevent spillage.11.Periodic observations of this system should be made to assure that urine is flowing freely.Volume on collection container has been calibrated without uro-prep¿ tray in place.Bard and uro-prep are trademarks and/or registered trademarks of c.R.Bard, inc.©2010 c.R.Bard, inc.All rights reserved.Directions for taking specimen: after catheter has been inserted into the bladder and an initial flow of urine is seen in the collection bag: 1.Clamp off catheter using white clamp.2.Insert plastic tube (inside bag) into the back of the sample device.3.Open sample device over the sample container.4.Unclamp catheter.5.Allow desired amount of urine to drain into collection container.6.Clamp off catheter.".
 
Event or Problem Description
It was reported that the catheters were missing from the catheter tray.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URETHRAL CATHETERIZATION TRAY WITH PLASTIC CATHETER
Common Device Name
URETHRAL CATHETER TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7993771
Report Number1018233-2018-04874
Device Sequence Number11090662
Product Code KOD
UDI-Device Identifier00801741030444
UDI-Public(01)00801741030444
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number772514
Device Catalogue Number772514
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/04/2018
Supplement Date Received by Manufacturer11/07/2018
Initial Report FDA Received Date10/22/2018
Supplement Report FDA Received Date12/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-