• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO S. DE R.L DE C.V. KYPHON XPANDER¿ INFLATABLE BONE TAMP; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO S. DE R.L DE C.V. KYPHON XPANDER¿ INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number K08A
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was pre-operatively diagnosed with osteoporotic l2 vertebral fracture and underwent balloon kyphoplasty (bkp) procedure.During surgery, while inflating the balloon, at 200 psi, diffusion of contrast medium was detected.Upon deflation and visual inspection, it revealed that there was a tear at the balloon.The product came in contact with the patient.No patient complications were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: product analysis: visual and optical analysis of the ibt revealed the balloon has ruptured/been cut at the center of the balloon when inflated.Functional check with a sample syringe confirmed the balloon was not able to inflate.This type of rupture is consistent with the balloon coming in contact with bone splinters while the balloon is inflated in the vertebral body.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER¿ INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L DE C.V.
avenida paseo cucapah # 10510
parque industrial ei lago
tijuana,bc 22570
MX  22570
MDR Report Key7993955
Report Number9612164-2018-02868
Device Sequence Number201668
Product Code HRX
UDI-Device Identifier00643169103054
UDI-Public00643169103054
Combination Product (Y/N)N
Initial Reporter CountryGR
PMA/510(K) Number
K041454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/10/2020
Device Catalogue NumberK08A
Device Lot Number0008919224
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Initial Date Received by Manufacturer 09/27/2018
Supplement Date Received by Manufacturer11/01/2018
12/19/2018
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date11/27/2018
12/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-