MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551PNAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551PNAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551PNAL
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Customer reported via phone call that insulin pump had rejected new battery.Customer¿s blood glucose was unknown at the time of incident.The customer also stated that the insulin pump buttons were harder to press.Troubleshooting was not performed.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device received with broken battery cap noted.All buttons functioning properly.No button error alarm during testing.No moisture or corroded keypad found.No flatted keypad.Connector was inspected and locked on lcd board.Device passed displacement test, a21 error test, self test and all operating currents tested within the specific range.No unexpected low battery alarm were noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551PNAL
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7994528
Report Number2032227-2018-47938
Device Sequence Number1190101
Product Code OZO
UDI-Device Identifier00643169507289
UDI-Public(01)00643169507289
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551PNAL
Device Catalogue NumberMMT-551PNAL
Device Lot NumberA3551PNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2019
Initial Date Received by Manufacturer 10/09/2018
Supplement Date Received by Manufacturer01/07/2020
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date01/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
Patient Weight56
-
-