| Catalog Number |
0066213 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The immucor laboratory tested retention product on 05oct2018, which performed as expected.The immucor laboratory also tested a returned blood sample from the customer site with retention product on 05oct2018, which yielded the expected positive outcomes, and as a result the customer's issue was not reproduced or confirmed.Immucor technical support used a remote electronic connection method on (b)(6) 2018 to assess the testing instrument.The immucor internal identification number for this report is (b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a european (b)(6) customer site reported, via email, an unexpected negative antibody screen when using capture-r ready-screen (pooled cells) with a galileo instrument.
|
| |
|