• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS HIGH FLEXION INSERT SZ 1-2 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS HIGH FLEXION INSERT SZ 1-2 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 71421501
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Injury (2348)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported from a literature review, that the patient returned to the hospital 14 months after the primary surgery.The insert was found dislocated and a revision surgery was performed due to this.
 
Additional Manufacturer Narrative
The associated complaint device was not returned.The clinical/medical team concluded, per the literature review with a single x-ray photo, the retained posterior osteophyte most likely led to impingement during flexion and eventual anterior dislocation of the insert while weight-bearing (down stairs) which was supported by the resolution of the clicking post-tka insert-only revision with full rom/flexion.No further patient information was provided.No additional impact to the patient is expected, as it was documented that the patient¿s recovery was uneventful, experienced an increase of flexion (by 10%), and the clicking sensation had disappeared.No further medical assessment is warranted at this time.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HIGH FLEXION INSERT SZ 1-2 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7994985
Report Number1020279-2018-02249
Device Sequence Number1190571
Product Code JWH
UDI-Device Identifier03596010492388
UDI-Public03596010492388
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K032295
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71421501
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/26/2018
Supplement Date Received by Manufacturer09/26/2018
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date01/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age76 YR
-
-