BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558390
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the common bile duct (cbd) during a lithotomy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the balloon was leaking liquid.Reportedly, the balloon was removed from the scope and the balloon was noted to have ruptured.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
(b)(4).A visual examination of the complaint device revealed that there were no visual defects.Functional analysis was performed, the balloon was inflated, and a leak was found due to a pinhole located at the proximal shoulder of the balloon.The most probable root cause is manufacturing deficiency, as the design or validation of a suppliers manufacturing process was not sufficient to ensure the finished device met the intent of the design.An investigation to address this issue has been completed.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the common bile duct (cbd) during a lithotomy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the balloon was leaking liquid.Reportedly, the balloon was removed from the scope and the balloon was noted to have ruptured.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE WIREGUIDED
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key7995976
Report Number3005099803-2018-60786
Device Sequence Number16215644
Product Code FDF
UDI-Device Identifier08714729339328
UDI-Public08714729339328
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110833
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/03/2020
Device Model NumberM00558390
Device Catalogue Number45735
Device Lot Number0021212808
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2018
Initial Date Received by Manufacturer 09/29/2018
Supplement Date Received by Manufacturer10/25/2018
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date11/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
Patient Weight55
-
-