• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS HIGH FLEXION INSERT SZ 1-2 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS HIGH FLEXION INSERT SZ 1-2 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 71421501
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Injury (2348)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported during a literature review, after first revision surgery from complaint (b)(4) , one month following the revision the patient developed a postoperative infection secondary to a (b)(4), which ultimately required a two-stage revision surgery.
 
Additional Manufacturer Narrative
The associated genesis ps high flexion insert was not returned for evaluation.This event was reported from a literature review.After repeated requests, smith and nephew has been unable to obtain device details.Smith and nephew has an outstanding request with the reporter for information.As device details were not made available, device history record review cannot be completed.Also, a complaint history review and sterilization documentation review could not be performed due to lack of device information.Our clinical evaluation noted that based on the article, the first revision was precipitated by the dislocation of the insert secondary to a fall from a bus.The root cause of the methicillin-resistant staphylococcus aureus that resulted in the second revision cannot be concluded but is not due to a failure of the device or procedure.With no supporting clinical/medical documents a thorough investigation cannot be performed.Without the return of the actual product involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HIGH FLEXION INSERT SZ 1-2 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7996288
Report Number1020279-2018-02255
Device Sequence Number13799559
Product Code JWH
UDI-Device Identifier03596010492388
UDI-Public03596010492388
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
K032295
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71421501
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/26/2018
Supplement Date Received by Manufacturer09/26/2018
Initial Report FDA Received Date10/23/2018
Supplement Report FDA Received Date04/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age60 YR
-
-