• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM PILLCAM; SYSTEM, IMAGING, GASTROINTESTINAL, WIRELESS, CAPSULE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM PILLCAM; SYSTEM, IMAGING, GASTROINTESTINAL, WIRELESS, CAPSULE Back to Search Results
Model Number FGS-0391
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Injury (2348)
Date of Event 09/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the doctor identified that the patient had a retained capsule when the data or study was reviewed, so an x-ray and ct scan was performed for confirmation of the capsule not passing to the cecum.The patient is still an inpatient and is on tube feeding.The patient is not experiencing any discomfort or pain from the capsule, but was operated to obtain the retained capsule.A laparotomy was performed, the capsule was removed and a partial resection of the ileum was undertaken.The capsule was located laparoscopically in the terminal ileum where it was caught by multiple benign strictures, caused by nonsteroidal anti-inflammatory drug (nsaid) overuse.The patient had returned home.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PILLCAM
Common Device Name
SYSTEM, IMAGING, GASTROINTESTINAL, WIRELESS, CAPSULE
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
MDR Report Key7996481
Report Number9710107-2018-01009
Device Sequence Number1381233
Product Code NEZ
UDI-Device Identifier07290101364085
UDI-Public7290101364085
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K123864
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 10/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2018
Device Model NumberFGS-0391
Device Catalogue NumberFGS-0391
Device Lot Number34618U
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2018
Initial Report FDA Received Date10/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
Patient Weight50
-
-